Pupillometer Consistency and Reproducibility

High reproducibility measurements for pupil size in both dim and bright light conditions were found, with Cronbach’s alpha values of 0.987 and
0.985, respectively.

Intraclass Correlation Coefficient (ICC) for average measures was 0.981 (95% CI: 0.959 to 0.991) for dim light and 0.973 (95% CI: 0.909 to 0.988)
for bright light.

For accommodation measurements, the results demonstrated good consistency, with a Cronbach’s alpha of 0.837 and an ICC for average
measures of 0.829 (95% CI: 0.724 to 0.898).

Pupillary reactivity was categorized into three groups based on constriction velocity: non-reactive (velocity < 0.4 mm/s), sluggish (0.4 mm/s ≤ velocity < 0.8 mm/s), and brisk (velocity ≥ 0.8 mm/s). The measured reproducibility among these categories was 100%.

Table 1. Reproducibility analysis of pupillary metrics in control subjects

MetricCronbach’s alphaICC – Average Measures95% CI (Avg)ICC – Single Measures95% CI (Single)
Dim Light (Diameter, mm)0.9870.9810.959–0.9910.9460.886–0.973
Bright Light (Diameter, mm)0.9850.9730.909–0.9880.9220.770–0.966
Reactive Velocity (mm/s)0.7160.7170.544–0.8320.4580.284–0.622
Pupil Reactivity Category1.0001.0001.000
Accommodation (mm)0.8370.8290.724–0.8980.6180.467–0.747

ICC = Intraclass Correlation Coefficient; CI = Confidence Interval